New Research Reports Low Rate of Mechanical Failure In Optim Family Leads During Long-Term Evaluation

First-of-its-kind study published in HeartRhythm assesses Optim-insulated leads in more than 10,000 patients over three years of follow-up

December 05, 2014

Media Contact

Shane Osborne
Heart Rhythm Society
media@hrsonline.org
202-464-3431

A new study evaluating Optim-insulated implantable cardioverter defibrillator (ICD) leads found low rates of all-cause mechanical failure during a median follow-up of 3.2 years. The long-term prospective study, published in the December edition of HeartRhythm,the official journal of the Heart Rhythm Society, is the first-of-its-kind to observe a large population of patients over a relatively long follow-up period. The study assessed Durata DF4, Durata DF1 and Riata ST Optim leads to determine the rates of all-cause mechanical failure and its subtypes, including conductor fracture, insulation abrasion and externalized conductors.

The study independently analyzed data from three prospective registries launched by St. Jude Medical between 2006 and 2009. The registries enrolled 10,835 patients (mean age of 66 years and a mean left ventricular ejection fraction of 29 percent) with 11,016 Optim-insulated leads implanted (8,147 Durata and 2,869 Riata ST Optim leads). All registries included standardized baseline documentation, six month follow-up, adverse events reports and documentation of lead revisions or inactivation, study withdrawal and death.

During a median follow-up of 3.2 years, there were 51 mechanical failures (0.46 percent), with 99 percent of patients free of this outcome by five years of follow-up. In addition, freedom from conductor fracture was identified in 99.4 percent of patients and all-cause abrasion in 99.8 percent of the leads, and there were no reports of externalized conductors.

The positive results come at a time when ICDs are implanted at a high rate of approximately 10,000 implants per month in the United States (U.S.)[i] . Additionally, the mechanical failures in older Riata leads led to class action lawsuits and Food and Drug Administration recalls left many people skeptical of their ability to be effective. The new Optim-insulated leads were designed to have greater durability and lower rates of abrasion than older designs.

“Previous studies showed positive results for Optim leads, but only followed patients for about two years and evaluated a significantly smaller number of patients with leads. Our ongoing analyses now include some patients followed for up to seven years and support existing data that better performance with Optim leads is a reality,” stated lead-author John A. Cairns, MD, Professor of Medicine, Division of Cardiology of the University of British Columbia in Vancouver, British Columbia. “The data from this study is of great importance to help electrophysiologists make more informed decisions as they relate to the performance of these leads and the management of patients who depend on them.”

Failure of high-voltage leads have major implications for patient morbidity and mortality and can be very costly for the U.S. health care system. This study is ongoing and will report on long-term follow-up of these patients and leads in the future. The study also highlights the need for prospective observational studies to be initiated at the time of introduction of a new lead to provide reliable data regarding lead survival.   

[i] ICD Registry Program. Retrieved November 25, 2014, from Heart Rhythm Society Web site: http://www.hrsonline.org/Practice-Guidance/Quality-Outcomes-Reporting/ICD-Registry#axzz3K6iLWlcY

About the Heart Rhythm Society

The Heart Rhythm Society is the international leader in science, education and advocacy for cardiac arrhythmia professionals and patients, and the primary information resource on heart rhythm disorders. Its mission is to improve the care of patients by promoting research, education and optimal health care policies and standards. Incorporated in 1979 and based in Washington, DC, it has a membership of more than 5,900 heart rhythm professionals in more than 70 countries around the world. For more information, visit www.HRSonline.org .

About the HeartRhythm Journal

HeartRhythm provides rapid publication of the most important science developments in the field of arrhythmias and cardiovascular electrophysiology (EP). As the official journal of the Heart Rhythm Society, HeartRhythm publishes both basic and clinical subject matter of scientific excellence devoted to the EP of the heart and blood vessels, as well as therapy. The journal is the only EP publication serving the entire electrophysiology community from basic to clinical academic researchers, private practitioners, technicians, industry and trainees. HeartRhythm has an impact factor of 4.866 (as of 2016) and ranks 23rd out of 114 cardiovascular medicine journals worldwide by the Institute for Scientific Information, remaining the number one specialty journal in cardiology. It is also the official publication of the Cardiac Electrophysiology Society.